ULVAC VACUUM > Gas analyzers > Gas analyzers > Residual Gas Analysis/Process Gas Monitor High Performance Process Gas Monitor Qulee HGM
Residual Gas Analysis/Process Gas Monitor High Performance Process Gas Monitor Qulee HGM
![]() |
||||
|
||||
|
ULVAC's new residual gas analyzer QULEE series, with revolutionary simplicity and usability, now offers a high-end model "QULEE HGM". Achieving highest sensitivity among all models at 2.5 x 10-6 A/Pa, QULEE HGM can meet the needs in various R&D and other vacuum system process monitoring.
- Achieved highest senstivity 2.5 x 10-6 A/Pa(Ie=1mA)
Max. 250ºC High temperature baking while connecting the sensor.
Display integrated design
Simple operation-"One Click" function
Total pressure measurement
Qulee QCS Software is included
CE MARKING
RoHS Compliance
- Various types of gas analysis for research and development.
Residual gas analysis for ultra high vacuum pumping equipment, etc.
Residual gas analysis in PVD system and vacuum deposition.
R&D for Photo voltaic field.
| Item | Qulee HGM | |||
| Model no. | HGM-202 | HGM-302 | ||
| Mass Filter Type | Quadrupole | |||
| Mass Range | 1-200 amu | 1-300 amu | Resolution | M/ΔM=1M(10% P.H.) |
| Detector Type | Faraday Cup | SEM | Faraday Cup | SEM |
| Sensitivity | 2.5 x 10-6 A/Pa | 4A/Pa | 2.0 x 10-6 A/Pa | 4A/Pa |
| Minimum Detectable Partial Pressure | 10-9Pa Level | 10-13Pa Level | 10-9Pa Level | 10-13Pa Level |
| Maximum Operating Pressure | 1x10-2Pa | |||
| Toatal Puresure Function | capable | |||
| One Click function | capable/He/H2O/N2/O2/Any Gas |
|||
| One Click Leak Test function | capable | |||
| Power Input | DC24V | |||
| Maximum Sensor Operating Temperature | 250ºC | |||
| Maximum Bakeout Temperature (electronics removed) |
300ºC | |||
| Operating Temperature | 10 - 40ºC | |||
| Standard Software | Qulee QCS (Windows 2000/XP/Vista) | |||
| CE MARKING | Compliance | |||